You are here: Experiment Guide > Imaging Modes > Electrochemistry > Math Functions

Math Functions

Three Math functions are provided:

To Program Math Functions

  1. Press the Math button.
  2. Press Enter to continue. Alternatively, pressing the Math button again clears (i.e., resets) the maximum and minimum readings recorded since the last reset.
NOTE: The stored maximum and minimum readings are cleared when the controller is turned off.
  1. Toggle the arrow buttons as needed to select Input A or Input B.
  2. Press Enter.
  3. Press Enter for the Max/Min function. Toggle the arrow buttons as needed to select among:
  4. Press Enter.
NOTE: This concludes this procedure unless you are also enabling Linear Equation and/or Filter functions. Proceed to step 4 to specify a linear equation. Proceed to step 10 to specify filtering.

To specify a linear equation:

  1. Ensure you have carried out the steps described under To Program Math Functions, above.
  2. Press Enter repeatedly until Linear Equ MX+B appears in the lower half of the LCD.
  3. Toggle the arrow buttons as needed to select MX+B or M(X+B), where M is the slope of the line, B is the y-axis offset in the first equation, and X is the sensor reading.
  4. Press Enter.
  5. Enter a numeric M value.
  6. Press Enter.
  7. Toggle the arrow buttons as needed to select among linear X variables:
  8. Press Enter.
  9. Toggle the arrow buttons as needed to select among linear B variables:
    NOTE: SP stands for setpoint.
  10. Press Enter.
  11. Enter a numeric B value.
  12. Press Enter.
NOTE: This concludes this procedure unless you are also enabling the Filter function. Proceed to step 10 to specify filtering.

To specify filtering:

  1. Ensure you have carried out the steps described under To Program Math Functions, above.
  2. Press Enter repeatedly until Filter On appears in the lower half of the LCD.
  3. Toggle the arrow buttons as needed to select Filter On or Filter Off.
NOTE: If you select Filter Off the controller returns to normal operation and you may exit this procedure.
  1. Toggle the arrow buttons as needed to select between 02 and 64 filter points. The default value is 08. Exponential filtering is applied to a moving average of the number of points selected.
  2. Press Enter.
  3. Toggle the arrow buttons as needed to select between a 01% (the default) and 10% filter window. If a single reading is greater than the filter window value, the reading is treated as intentional and the filter is restarted. Percentages are calculated over the full scale range.

 

www.bruker.com Bruker Corporation
www.brukerafmprobes.com 112 Robin Hill Rd.
nanoscaleworld.bruker-axs.com/nanoscaleworld/ Santa Barbara, CA 93117
   
  Customer Support: (800) 873-9750
  Copyright 2010, 2011. All Rights Reserved.